Recall Z-1101-2021 — GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column
U.S. FDA — Device Enforcement · United States · 2020-12-10 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.