Recall Z-1045-2014 — CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES A
U.S. FDA — Device Enforcement · United States · 2014-02-12 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.