Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0826-2024 — Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;

U.S. FDA — Device Enforcement · United States · 2023-11-29 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0826-2024
Date of alert
2023-11-29
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
American Contract Systems, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00191072117038
Reported risk: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.