Recall Z-0766-2020 — Apollo EZ, Model Number 9784152036 - Product Usage: The Apollo EZ remote-controlled table is X-ray equipment that can be used to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-contr
U.S. FDA — Device Enforcement · Italy · 2019-11-04 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.