Recall Z-0763-2019 — Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan
U.S. FDA — Device Enforcement · United States · 2018-12-21 · Class I
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.