Recall Z-0753-2025 — The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.
U.S. FDA — Device Enforcement · United States · 2024-11-21 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.