Recall Z-0499-2019 — Dimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista¿ System.Cat. No. K2054, Mat. No. 10464323
U.S. FDA — Device Enforcement · United States · 2018-10-11 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.