Recall Z-0496-2013 — SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. implantable infusion pump. The pump is part of an implanted infusion system that consists of a SynchroMed EL pump and a catheter. The SynchroMed EL Infusion System is indicated for use when patient t
U.S. FDA — Device Enforcement · United States · 2012-11-09 · Class I
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.