Reported risk: Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.