Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0472-2023 — ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

U.S. FDA — Device Enforcement · United States · 2022-11-09 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0472-2023
Date of alert
2022-11-09
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Siemens Medical Solutions USA, Inc
Product category
device
Product type
Devices
Reported risk: When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.