Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0399-2021 — Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured from stainless steel. To assist in inserting catheters, introducer sheaths, and other medical devices in vascular applications.

U.S. FDA — Device Enforcement · United States · 2020-10-22 · Class I

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0399-2021
Date of alert
2020-10-22
Country / jurisdiction
United States
Risk level
Class I
Risk type
recall
Brand
Cook Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
13199931
Reported risk: Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.