Recall Z-0290-2019 — Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of
U.S. FDA — Device Enforcement · Switzerland · 2018-08-28 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.