Recall Z-0279-2013 — Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.
U.S. FDA — Device Enforcement · United States · 2012-10-16 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.