Recall Z-0268-2020 — VITROS chemistry Products Cl- Slides, Catalog No. 8445207 (which does NOT support Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.
U.S. FDA — Device Enforcement · United States · 2019-09-10 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.