Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0257-2016 — GE Healthcare, SIGNA HD 3.0T. Product Usage: The Signa Profile EXCITE MR system is an open, whole body scanner for use as a diagnostic imaging device. It may be utilized for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR safe biopsy

U.S. FDA — Device Enforcement · United States · 2015-08-31 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0257-2016
Date of alert
2015-08-31
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
GE Healthcare
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00000000036657
Reported risk: When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.