Recall Z-0228-2026 — LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
U.S. FDA — Device Enforcement · United States · 2025-09-26 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.