Recall Z-0222-2025 — CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0
U.S. FDA — Device Enforcement · United States · 2024-07-16 · Class I
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.