Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0213-2022 — Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20,

U.S. FDA — Device Enforcement · United States · 2021-09-24 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0213-2022
Date of alert
2021-09-24
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Philips North America Llc
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00884838099258
Reported risk: When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.