Recall Z-0207-2019 — Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a non-sterile condition. The inserter is designed to enter the disc space, with a fixed engagement between the inserter and the cage and th
U.S. FDA — Device Enforcement · United States · 2018-09-12 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.