Recall Z-0203-2019 — IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that can receive information from a hospital computer system (also called a Laboratory Information System [LIS]).
U.S. FDA — Device Enforcement · United States · 2018-08-29 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.