Recall Z-0095-2019 — GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CA

U.S. FDA — Device Enforcement · Finland · 2018-09-04 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0095-2019
Date of alert
2018-09-04
Country / jurisdiction
Finland
Risk level
Class II
Risk type
recall
Brand
GE Healthcare Finland Oy
Product category
device
Product type
Devices
Reported risk: When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.