Recall Z-0094-2019 — Widex BABY hearing aid (referred to as BABY440), powered by zinc-air batteries size 10 Intended use: BABY440 is intended for, but not limited to, babies (0 to 3 years of age) to be used as air conduction amplification devices in everyday listening environments.
U.S. FDA — Device Enforcement · Denmark · 2018-08-31 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: The reason for this field action is due to the mounting of a non-tamper-resistant (ordinary) battery drawer on some BABY440 hearing aids. The correct battery drawer is tamper-resistant and require a special tool to open it.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.