Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-0089-2026 — Model Number L211 PROPONENT DR SL MRI Pacemaker

U.S. FDA — Device Enforcement · United States · 2025-08-20 · Class I

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0089-2026
Date of alert
2025-08-20
Country / jurisdiction
United States
Risk level
Class I
Risk type
recall
Brand
Boston Scientific Corporation
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00802526559105
Reported risk: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.