Baxter Clearlink System Buretrol Solution Set, 115" (2.9 m) with 150 mL Clearlink Burette (Ball-Valve Drip Chamber) (Non-DEHP); a sterile Rx IV fluid pathway device Product Usage: The device is used for the administration of fluids from a container into the patient's vascular system through a va — recall Z-0015-2013
U.S. FDA — Device Enforcement · United States · 2012-09-07 · Class I
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.