Product Wiki · Recalls · U.S. CPSC

Delta Gas Ball Valves — recall 25255

U.S. CPSC · United States · 2025-05-08

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. CPSC
Recall identifier
25255
Date of alert
2025-05-08
Country / jurisdiction
United States
Risk type
recall
Brand
Masco Canada Limited, of Canada
Product category
consumer_product
Reported risk: This recall involves all models of Delta-branded Gas Ball Valves. These are chrome manually operated gas valves sold by Masco Canada Limited and Delta Faucet Company for use in school, research and hospital laboratories. The valves are used to control the flow of gas, acting as an on/off valve for gas delivery. All of the models will have the word "Delta" on at least one side of the chrome-colored sides of the body and the word "GAS" written in white on the blue-colored top. The recalled models are also identifiable by the "1/2 PSI MAX" and "GAS/GAZ" stamped on the bottom. Hazards: Small amounts of gas can leak from the valve when it is in the open position, posing fire and burn hazards.
Primary source: https://www.cpsc.gov/Recalls/2025/Delta-Gas-Ball-Valves-Recalled-Due-to-Fire-and

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.