Product Wiki · Recalls · Australia ACCC Product Safety

Recall all-minimed-640g-670g-700-770g-and-780g-insulin-pumps — All MiniMed 640G, 670G, 700, 770G, and 780G Insulin Pumps

Australia ACCC Product Safety · AU · 2022-06-09 · Product Recall

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
Australia ACCC Product Safety
Recall identifier
all-minimed-640g-670g-700-770g-and-780g-insulin-pumps
Date of alert
2022-06-09
Country / jurisdiction
AU
Risk level
Product Recall
Risk type
product_recall
Brand
All MiniMed
Product category
Pharmacy and medical products
Reported risk: The gold battery cap contact may fall off when the user is charging the battery. If the contact falls off, the 'Insert battery' alarm will not clear automatically when the user inserts a new AA battery as the circuit is not complete. This will result in the alarm escalating to a siren and then the pump will shut down and will not power up.
Primary source: https://www.productsafety.gov.au/search-consumer-product-recalls/all-minimed-640

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.